Optimization and validation of a high-throughput nanofluidic real-time PCR assay to evaluate nasopharyngeal carriage of 15 bacterial species and 92 Streptococcus pneumoniae serotypes

Article scientifique 2022 Sciences de la santé Anglais

Résumé

Abstract Background: Sensitive tools for detecting concurrent colonizing pneumococcal serotypes are needed for detailed evaluation of the direct and indirect impact of routine pneumococcal conjugate vaccine (PCV) immunization. Method: A high-throughput quantitative nanofluidic real-time PCR (Fluidigm) reaction-set was developed to detect and quantify 92 pneumococcal serotypes in archived clinical samples. Nasopharyngeal swabs collected in 2009-2011 from South African children ≤5years-old, previously serotyped with standard culture-based methods were used for comparison. Results: The reaction-set within the Fluidigm® effectively amplified all targets with high efficiency (90-110%), reproducibility (R2≥0.98), and at low limit-of-detection (<102 CFU/ml). A blind analysis of 1973 nasopharyngeal swab samples showed diagnostic sensitivity >80% and specificity >95 compared with the referent standard, culture-based Quellung method. The Fluidigm method was able to serotype pneumococcal types with good discrimination compared with Quellung (ROC-AUC: >0.73). Conclusion: The high-throughput nanofluidic real-time PCR method simultaneously detects 57 individual serotypes, and 35 serotypes within 16 serogroups in 96 samples (including controls), within a single qPCR run. This method can be used to evaluate the impact of current PCV formulations on vaccine-serotype and non-vaccine-serotype colonization, including detection of multiple concurrently colonizing serotypes. Interpretation: The Fluidigm method can allow for monitoring of serotype-specific bacterial load, as well as emergence or ongoing transmission of minor or co-colonizing serotypes that may have invasive disease potential.

Citer ce document

Downs, S., Madhi, S. A., Merwe, L. V. D., Nunes, M. C., & Olwagen, C. P. (2022). Optimization and validation of a high-throughput nanofluidic real-time PCR assay to evaluate nasopharyngeal carriage of 15 bacterial species and 92 Streptococcus pneumoniae serotypes. Research Square. https://doi.org/10.21203/rs.3.rs-2335378/v1

Exporter : BibTeX · RIS (Zotero, Mendeley, EndNote)

Accès au document

Texte intégral archivé sur SPHAERO, en accès ouvert : lisible gratuitement en ligne après connexion. Le document reste aussi disponible à sa source. Se connecter

Voir l'article sur le site de la revue

Licence et provenance

Licence : CC BY

Notice moissonnée depuis OpenAlex le 10/09/2026. Le document reste hébergé par sa source.
Voir le document à la source →

Statistiques

Consultations (hors robots) : 14

Téléchargements (hors robots) : 0

Méthode de comptage