{# Audit 04/10/2026 : « autre » n'est pas un code de langue ; SPHAERO n'est pas l'éditeur des documents qu'elle héberge ou référence. #} {# citation_pdf_url doit mener à un PDF : un lien vers une page DOI est pénalisé par Google Scholar (avant : tout lien externe). #}
Accès ouvert · CC BY

Micellar eco-friendly HPLC method for simultaneous analysis of ternary combination of aspirin, atorvastatin and ramipril: application to content uniformity testing

Article scientifique 2023 Anglais

Résumé

BACKGROUND: Cardiovascular disease medications such as aspirin (ASP), statins like atorvastatin (ATR), and blood pressure-lowering drugs including ACE inhibitors like ramipril (RAM) have been included in the World Health Organization (WHO) Essential Medicines List (EML) for many years. Therefore, there is a strong demand to develop a simple, rapid, and sensitive analytical method that can detect and quantitate the ternary mixture of these analytes in pharmaceutical preparations in a short run time. Lately, the analytical community focused on eliminating or reducing hazardous chemicals and solvents usage. RESULTS: A green, fast, selective, and cost-effective micellar HPLC method was established and validated for the concurrent determination of ternary combination of ASP, ATR, and RAM in the pure form and pharmaceutical preparations. Resolution of the three drugs was achieved by using a monolithic column and a micellar mobile phase consists of 0.3% triethylamine (TEA) in 90: 10 an aqueous solution of 0.12 M sodium dodecyl sulfate (SDS): n-propanol, (v/v). The pH was adjusted to 2.5 using orthophosphoric acid and a flow rate of 1.5 mL/min. was applied. To ensure method reproducibility, Valsartan (VAL) was utilized as an internal standard (IS). The UV detection of the studied drugs was performed at 210 nm. Good linearity for the three drugs was obtained over the concentration ranges of 1.0-200.0 mg/mL, 0.5-200.0 mg/mL, and 5.0-100.0 mg/mL with correlation coefficients of 0.9998,0.9999 and 0.9999 for ASP, ATR, and RAM respectively. The method sensitivity was revealed by the relatively small values of limits of detection (LOD) (0.19, 0.13 and 0.30 mg/mL) and limits of quantitation (LOQ) (0.63, 0.44 and 0.99 mg/mL) for ASP, ATR, and RAM, respectively. The retention times of ASP, ATR and RAM were 1.50, 2.3 and 4.3 min., respectively. CONCLUSIONS: The suggested technique was employed for the analysis of the three drugs in their prepared tablets maintaining the recommended pharmaceutical ratio without any interference from excipients. The method was further extended to content uniformity testing of RAM. The results were validated according to international council for harmonisation (ICH) guidelines.

Citer ce document

Abdallah, N. A., El-Brashy, A. M., Ibrahim, F., & El‐Awady, M. I. (2023). Micellar eco-friendly HPLC method for simultaneous analysis of ternary combination of aspirin, atorvastatin and ramipril: application to content uniformity testing. BMC Chemistry. https://doi.org/10.1186/s13065-023-00929-y

Exporter : BibTeX · RIS (Zotero, Mendeley, EndNote)

Accès au document

Texte intégral en lecture en ligne, réservé aux abonnés SPHAERO et aux membres de l'institution. Se connecter

Voir l'article sur le site de la revue

Licence et provenance

Licence : CC BY

Notice moissonnée depuis OpenAlex le 16/09/2026. Le document reste hébergé par sa source.
Voir le document à la source →

Statistiques

Consultations : 2

Téléchargements : 0