Evaluation of GeneXpert PA assay compared to genomic and (semi-)quantitative culture methods for direct detection of Pseudomonas aeruginosa in endotracheal aspirates

Article scientifique 2021 Sciences de la santé Anglais

Résumé

INTRODUCTION: Pseudomonas aeruginosa is a common cause of ventilator-associated pneumonia (VAP). Rapid and accurate detection of lower respiratory tract colonization and/or infection with P. aeruginosa may advise targeted preventive (antibody-based) strategies and antibiotic therapy. To investigate this, we compared semi-quantitative culture results from 80 endotracheal aspirates (ETA) collected from mechanically-ventilated patients, to two culture and two non-culture-based methods for detection of P. aeruginosa. METHODS: P. aeruginosa-positive (n = 40) and -negative (n = 40) ETAs from mechanically ventilated patients analyzed initally by (i) routine semi-quantitative culture, were further analyzed with (ii) quantitative culture on chromogenic ChromID P. aeruginosa and blood agar; (iii) enrichment in brain heart infusion broth followed by plating on blood agar and ChromID P. aeruginosa; (iv) O-antigen acetylase gene-based TaqMan qPCR; and (v) GeneXpert PA PCR assay. RESULTS: Of the 80 ETA samples included, one sample that was negative for P. aeruginosa by semi-quantitative culture was found to be positive by the other four methods, and was included in an "extended" gold standard panel. Based on this extended gold standard, both semi-quantitative culture and the GeneXpert PA assay showed 97.6% sensitivity and 100% specificity. The quantitative culture, enrichment culture and O-antigen acetylase gene-based TaqMan qPCR had a sensitivity of 97.6%, 89.5%, 92.7%, and a specificity of 97.4%, 100%, and 71.1%, respectively. CONCLUSION: This first evaluation of the GeneXpert PA assay with ETA samples found it to be as sensitive and specific as the routine, hospital-based semi-quantitative culture method. Additionally, the GeneXpert PA assay is easy to perform (hands-on time ≈ 5 min) and rapid (≈ 55 min assay time). The combination of the high sensitivity and high specificity together with the rapid acquisition of results makes the GeneXpert PA assay a highly recommended screening technique. Where this equipment is not available, semi-quantitative culture remains the most sensitive of the culture methods evaluated here for P. aeruginosa detection in ETA samples.

Citer ce document

Schalk, T. V. D., Coppens, J., Timbermont, L., Turlej-Rogacka, A., Heirstraeten, L. V., Berkell, M., Yu, L., Lammens, C., Xavier, B. B., Matheeussen, V., Ieven, M., McCarthy, M., Jorens, P. G., Ruzin, A., Esser, M. T., Kumar‐Singh, S., Goossens, H., & Malhotra‐Kumar, S. (2021). Evaluation of GeneXpert PA assay compared to genomic and (semi-)quantitative culture methods for direct detection of Pseudomonas aeruginosa in endotracheal aspirates. Antimicrobial Resistance and Infection Control. https://doi.org/10.1186/s13756-021-00978-9

Exporter : BibTeX · RIS (Zotero, Mendeley, EndNote)

Accès au document

Texte intégral archivé sur SPHAERO, en accès ouvert : lisible gratuitement en ligne après connexion. Le document reste aussi disponible à sa source. Se connecter

Voir l'article sur le site de la revue

Licence et provenance

Licence : CC BY

Notice moissonnée depuis OpenAlex le 05/09/2026. Le document reste hébergé par sa source.
Voir le document à la source →

Statistiques

Consultations (hors robots) : 16

Téléchargements (hors robots) : 0

Méthode de comptage