{# Audit 04/10/2026 : « autre » n'est pas un code de langue ; SPHAERO n'est pas l'éditeur des documents qu'elle héberge ou référence. #} {# citation_pdf_url doit mener à un PDF : un lien vers une page DOI est pénalisé par Google Scholar (avant : tout lien externe). #}
Accès ouvert

comparison of the performance and energy resolution of CdTe and Si detectors in the X-ray Fluorescence studies of metal samples and Alloys

Article scientifique 2022 Anglais

Résumé

X-ray fluorescence provides a powerful means of non-destructively determining the elemental composition of a sample. X-rays from a Molybdenum (Mo) source was fired on copper, molybdenum, lead, steel and brass samples to determine their composition and relative abundance of their constituent elements. Two different detectors: the Cadmium Telluride (CdTe) and Silicon(Si) detectors were used to pick up the signals from the scattering of the X-rays at the sample surfaces and their energy resolutions as well as efficiencies were compared. With a non-noisy amplifier, the Si detector had a higher resolution (0.27 % ) when compared to the 0.38 % for the CdTe detector but it had a lower efficiency when compared to that of the CdTe detector. It was also discovered that higher energies produced lower detector efficiencies.

Citer ce document

Etim, I., Mfon, R. (2022). comparison of the performance and energy resolution of CdTe and Si detectors in the X-ray Fluorescence studies of metal samples and Alloys. https://doi.org/10.46481/jnsps.2022.926

Accès au document

Texte intégral en lecture en ligne, réservé aux abonnés SPHAERO et aux membres de l'institution. Se connecter

Voir l'article sur le site de la revue

Statistiques

Consultations : 3

Téléchargements : 0