Performance evaluation of the touchscreen-based Muse™ Auto CD4/CD4% single-platform system for CD4 T cell numeration in absolute number and in percentage using blood samples from children and adult patients living in the Central African Republic

Article scientifique 2016 Sciences de la santé Anglais

Résumé

BACKGROUND: The new microcapillary and fluorescence-based EC IVD-qualified Muse™ Auto CD4/CD4% single-platform assay (EMD Millipore Corporation, Merck Life Sciences, KGaA, Darmstadt, Germany) for CD4 T cell numeration in absolute number and in percentage was evaluated using Central African patients' samples compared against the reference EC IVD-qualified BD FACSCount (Becton-Dickinson, USA) flow cytometer. METHODS: EDTA-blood samples from 124 adults, 10 adolescents, 13 children and 3 infants were tested in parallel at 2 reference laboratories in Bangui. RESULTS: = 0.99 and 0.98, respectively). The mean absolute bias between Muse™ and BD FACSCount cells in absolute number and percentage were -5.91 cells/µl (95% CI -20.90 to 9.08) with limits of agreement from -77.50 to 202.40 cells/µl, and +1.69 %CD4 (95% CI ±1.29 to +2.09), respectively. The percentages of outliers outside the limits of agreement were nearly similar in absolute number (8%) and percentage (10%). CD4 T cell counting by Muse™ allowed identifying the majority of individuals with CD4 T cell <200, <350 or <750 cells/µl corresponding to the relevant thresholds of therapeutic care, with sensitivities of 95.5-100% and specificities of 83.9-100%. CONCLUSIONS: The Muse™ Auto CD4/CD4% Assay analyzer is a reliable alternative flow cytometer for CD4 T lymphocyte enumeration to be used in routine immunological monitoring according to World Health Organization recommendations in HIV-infected adults as well as children living in resource-constrained settings.

Citer ce document

Mossoro‐Kpinde, C. D., Kouabosso, A., Bouassa, R. M., Longo, J. D. D., Kokanzo, E., Féissona, R., Grésenguet, G., & Bélec, L. (2016). Performance evaluation of the touchscreen-based Muse™ Auto CD4/CD4% single-platform system for CD4 T cell numeration in absolute number and in percentage using blood samples from children and adult patients living in the Central African Republic. https://doi.org/10.1186/s12967-016-1082-7

Exporter : BibTeX · RIS (Zotero, Mendeley, EndNote)

Accès au document

Texte intégral archivé sur SPHAERO, en accès ouvert : lisible gratuitement en ligne après connexion. Le document reste aussi disponible à sa source. Se connecter

Voir l'article sur le site de la revue

Licence et provenance

Licence : CC BY

Notice moissonnée depuis OpenAlex le 25/08/2026. Le document reste hébergé par sa source.
Voir le document à la source →

Statistiques

Consultations (hors robots) : 11

Téléchargements (hors robots) : 0

Méthode de comptage