A multi-site laboratory evaluation of the MEDSCAN application for automated POC-CCA interpretation
Résumé
Introduction: Control efforts against schistosomiasis are hampered by the subjective interpretation of the point-of-care circulating cathodic antigen (POC-CCA) urine test, which limits diagnostic consistency. We developed MEDSCAN (Mobile-Enabled Diagnostics for Schistosomiasis Control Analytics), a mobile application that uses smartphone imaging and computer vision to automate POC-CCA interpretation. Methods: In a multi-site laboratory evaluation across the USA, the Netherlands, and Kenya, we compared MEDSCAN to visual G-Score interpretation and a benchtop lateral flow reader (LFR). Results: All three methods produced clear concentration-response relationships, with normalized machine-based metrics achieving AUROC ≥ 0.90. MEDSCAN demonstrated excellent inter-user reproducibility (intra-class correlation coefficients exceeding 0.94) and substantial agreement with both visual and LFR interpretations across sites. Some device-to-device variability was observed, attributable to differences in smartphone camera hardware and image processing; however, binary diagnostic outcomes remained robust across a heterogeneous set of smartphones. Discussion: These results establish operational positivity thresholds for MEDSCAN-based on test-line signal alone or normalized metrics-suitable for direct implementation in field surveillance workflows. A large-scale field study is underway to evaluate MEDSCAN under routine POC-CCA surveillance conditions in endemic settings.
Citer ce document
Exporter : BibTeX · RIS (Zotero, Mendeley, EndNote)
Accès au document
Texte intégral archivé sur SPHAERO, en accès ouvert : lisible gratuitement en ligne après connexion. Le document reste aussi disponible à sa source. Se connecter
Voir l'article sur le site de la revueLicence et provenance
Licence : CC BY
Notice moissonnée depuis OpenAlex le 01/10/2026. Le document reste hébergé par sa source.
Voir le document à la source →